• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/16 HOLE/336MM/RIGHT; CONDYLAR PLATE FIXATION IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 4.5MM VA-LCP CURVED CONDYLAR PLATE/16 HOLE/336MM/RIGHT; CONDYLAR PLATE FIXATION IMPLANT Back to Search Results
Model Number 02.124.416
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Injury (2348)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on (b)(6) 2020, a variable angle (va) distal femur plate was scheduled to be removed for an unknown reason.It is unknown if the implant was removed.Concomitant device reported: unknown screws (part # unknown, lot # unknown, quantity unknown), unknown plate (part # unknown, lot # unknown, quantity unknown).This report is for one (1) 4.5mm va-lcp curved condylar plate/16 hole/336mm/right this is report 1 of 1 for (b)(4).
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Additional event information: it was reported that the plate and screws were successfully removed.The patient had fell and re fractured leading to perform a knee revision.
 
Additional Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Customer quality investigation: the implant(s) was not returned, and the investigation will be completed based on the supplied image(s) from the attachment(s) located in notes & attachments section of the product complaint.The image(s) was reviewed, and no issues were noted with the implant(s) that would contribute to the adverse event and therefore is unconfirmed.As the implant(s) was not returned an as received, dimensional, material or drawing reviews are not applicable.A manufacturing record evaluation could not be performed, as the lot number could not be determined from the images provided.There is no indication that a design or manufacturing issue contributed to the complaint as it was unconfirmed.No new malfunctions were observed during this investigation (based on the images) that may have contributed to the complaint condition.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
4.5MM VA-LCP CURVED CONDYLAR PLATE/16 HOLE/336MM/RIGHT
Common Device Name
CONDYLAR PLATE FIXATION IMPLANT
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key10314545
Report Number2939274-2020-03278
Device Sequence Number16718737
Product Code JDP
UDI-Device Identifier10886982042934
UDI-Public(01)10886982042934
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K110354
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number02.124.416
Device Catalogue Number02.124.416
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/03/2020
Supplement Date Received by Manufacturer07/23/2020
08/03/2020
Initial Report FDA Received Date07/23/2020
Supplement Report FDA Received Date07/27/2020
08/04/2020
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - PLATES; UNK - SCREWS: TRAUMA
Outcome Attributed to Adverse Event Required Intervention;
Patient Age65 YR
Patient Weight99
-
-