H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, complaint and lot history, applicable previous investigation(s), labeling, applicable manufacturing records, and applicable fmea documents.Based on a review of this information, the following was concluded: the complaint of damaged safety mechanism was confirmed but the exact cause remains unknown.One 20 ga x 0.75 in miniloc infusion set was returned for evaluation.The orange component from the safety mechanism was returned detached from the winged needle housing.No apparent material damage was visible on the plastic components of the device.The needle was observed to be bent; however, it is unknown if the bend occurred during or after use.The safety mechanism could not be activated without proper assembly of the components.The orange component was able to be reinserted in the safety mechanism housing which allowed proper functionality of the safety mechanism.The condition of the device prior to use is unknown, the exact cause could not be determined.A lot history review (lhr) of asdqs0010 showed no other similar product complaint(s) from this lot number.
|