| Model Number |
20 |
| Medical Device Problem Code |
Failure to Power Up (1476)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Physio-control continues to investigate the reported failure and will submit a supplemental report on this event to the fda as provided by 21 cfr 803.56.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device's on/off button is not responsive.It requires a large amount of force to activate and sometimes it still doesn't turn the device on.As a result, defibrillation therapy may be delayed or unavailable, if needed.There was no report of patient use associated with the reported event.
|
| |
|
Additional Manufacturer Narrative
|
|
Physio-control evaluated the customer's device and verified the reported issue.Physio replaced the device's user interface pcb assembly to resolve the reported issue.After completing other unrelated repairs, proper device operation was observed through functional and performance testing.The device was returned to the customer for use.Physio-control further evaluated the removed user interface pcb assembly and determined that the cause of the reported issue was due to a worn on/off button.
|
| |
|
Event or Problem Description
|
|
The customer contacted physio-control to report that their device's on/off button is not responsive.It requires a large amount of force to activate and sometimes it still doesn't turn the device on.As a result, defibrillation therapy may be delayed or unavailable, if needed.There was no report of patient use associated with the reported event.
|
| |
|