BIOTRONIK AG, BUELACH, SWITZERLAND PANTERA LEO 3.25/12; BASIC CORONARY ANGIOPLASTY BALLOON CATHETER Back to Search Results
Model Number 367007
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/09/2020
Type of Reportable Event Malfunction
Event or Problem Description
The pantera leo balloon catheter was chosen for the treatment of a high calcified lesion in the rca.The pantera leo was inflated for 10 times at rbp inside the lesion.After deflation the balloon was pushed to the distal side and contrast medium was leaking.Therefore another balloon catheter was used for the procedure.
 
Additional Manufacturer Narrative
The returned instrument was subjected to a technical analysis and the corresponding production documentation was reviewed to establish whether a deviation from the manufacturing process could be the cause for the reported event.The technical investigation of the returned device revealed two tears in the balloon.The tear at the distal x-marker has a length of about 1/3 mm and the tear between the proximal x-ray marker and the balloon center has a length of about 5 mm.Scratches were observed in the balloon material in close vicinity to both tears.It seems likely that the damage of the balloon surface was caused by a hard, sharp-edged object pressing against the balloon from the outside.Review of the production documentation confirmed that the instrument was manufactured according to specifications and passed all in-process and final inspections.In addition to visual inspections each instrument is tested for air tightness by means of a helium leak test.We can therefore confirm that the instrument was delivered in a leak-proof condition.Based on the conducted investigations, no manufacturing or material related root cause could be determined.The root cause is most likely related to the patients anatomy (i.E.Severe calcification).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PANTERA LEO 3.25/12
Common Device Name
BASIC CORONARY ANGIOPLASTY BALLOON CATHETER
Manufacturer (Section D)
BIOTRONIK AG, BUELACH, SWITZERLAND
ackerstrasse 6
buelach CH-81 80
CH  CH-8180
MDR Report Key10315764
Report Number1028232-2020-03044
Device Sequence Number1044999
Product Code LOX
UDI-Device Identifier07640130415652
UDI-Public07640130415652
Combination Product (Y/N)N
Initial Reporter CountryCH
PMA/510(K) Number
K163660
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 07/21/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/28/2022
Device Model Number367007
Device Catalogue NumberSEE MODEL NO.
Device Lot Number03195363
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/21/2020
Initial Date Received by Manufacturer 07/21/2020
Supplement Date Received by Manufacturer09/08/2020
Initial Report FDA Received Date07/23/2020
Supplement Report FDA Received Date09/09/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-