Device has been returned for evaluation.
The customer¿s complaint of front panel flashing was confirmed.
The service technician observed the issue was due to a faulty dp board.
Unit was functioning normally with our test dp board.
In addition, a worn out video connector was causing poor connection, pc card reader was damaged, and pip cable was corroded.
Parts were replaced.
Minor wear on front panel and rear panel.
Unit passed electrical leak test and all functional tests.
All video output signals and images tested okay.
Unit has up to date nbi and software version.
The cause can be attributed to an electrical component failure.
No further information was reported.
|