| Model Number |
SOFTECHDO+20.0 |
| Medical Device Problem Code |
Device Handling Problem (3265)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
05/20/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
No discrepancies were found as it relates to the batch documentation analysis.The audit determined that all procedures in the manufacturing and packaging of the device has been carried out correctly.Batch reconciliation was 100%.The lens was inspected under a stereoscopic microscope at 10x magnification.The visual examination carried out showed the lens was free from any tears.The cartridge and lens are compatible for implantation.
|
| |
|
Event or Problem Description
|
|
Lenstec, inc.Received an email notification stating "lens torn during procedure.Incision had to be made bigger and doc had to use suture.Like lens implanted.Cart45s cartridge used.".
|
| |
|