|
It was reported that there was resistance when attempting to inflate the balloon of the catheter with water.Supposedly, the hub of the inflated part of the catheter kept filling instead of filling the balloon at the end of the catheter inside the patient.Therefore, the catheter was removed from the patient and a new 18fr foley catheter was inserted with no issues.
|
|
The reported event was inconclusive, as no sample returned for evaluation.A potential root cause for this failure mode could be due to pinch lumen/collapse lumen/thick sac thickness/valve damage/short inflation lumen.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "warning: do not use ointments or lubricants having a petrolatum base.They will damage latex.Visually inspect the product for any imperfections or surface deterioration prior to use.Use luer tip syringe to inflate with stated ml of sterile water.Or for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.For urological use only.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.This is a single use device.Do not resterilize any portion of this device.Reuse and/or repackaging may create a risk of patient or user infection, compromise the structural integrity and/or essential material and design characteristics of the device, which may lead to device failure, and/or lead to injury, illness or death of the patient." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the device was not returned.
|
|
It was reported that there was a resistance when attempting to inflate the catheter balloon with water.Supposedly, the hub of the inflated part of the catheter kept filling instead of filling the balloon at the end of the catheter inside the patient.Therefore, the catheter was removed from the patient and a new 18fr foley catheter was inserted with no issues.
|