MEDTRONIC PUERTO RICO OPERATIONS CO. SURESCAN; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
|
Back to Search Results |
|
Model Number 97714 |
Device Problems
Intermittent Continuity (1121); Therapy Delivered to Incorrect Body Area (1508); Patient Device Interaction Problem (4001)
|
Patient Problems
Undesired Nerve Stimulation (1980); Therapeutic Effects, Unexpected (2099)
|
Event Date 03/20/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 97740, serial#: (b)(4) , product type: programmer, patient.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a consumer regarding a patient with an implantable neurostimulator (ins) for spinal pain.It was reported that the caller stated that the patient had a hard time communicating with the patient programmer.The patient received a poor communication screen.The patient used the antenna with the patient programmer and attempted to connect without the antenna, but also received poor communication.It was also reported that the patient sat down to use the bathroom and they felt their stimulation surge through their lower stomach, so the patient was trying to turn stimulation down to use the restroom.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from patient who reported that the circumstances that led to stimulation in lower stomach is due to a sudden movement.Patient noted that all is well now with the new programmer for now.
|
|
Event Description
|
Additional information was received from a patient.It was reported that the patient received the new patient programmer (pp) and it only worked intermittently.During the call, the patient attempted to connect to his ins without the pp antenna and the patient received poor comm.His ins was charged and his stimulation was turned on.The patient had not any falls/traumas and the patient used alkaline batteries in the pp.It was very intermittent poor communication with and without antenna.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|