| Model Number |
ZCU300 |
| Medical Device Problem Code |
Adverse Event Without Identified Device or Use Problem (2993)
|
| Health Effect - Clinical Codes |
Blurred Vision (2137); No Code Available (3191)
|
| Date of Event |
02/06/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
(b)(4) all pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Event or Problem Description
|
|
It was reported that a tecnis toric ii intraocular lens (iol) was explanted from the patient's operative eye due to blurry with residual astigmatism.The symptom was observed at 1 day post-op exam.The patient's pre-op visual acuity was 20/40 and remains the same post surgery.There was no additional surgical intervention performed during the secondary procedure.No further information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
Section d10: device available for evaluation: yes.Section d10: returned to manufacturer on: 7/29/2020.Section h3: device returned to manufacturer: yes.Device evaluation: the product was received inside a non-jnj lens case.Additionally, a complaint intake form was received as well.Visual inspection under magnification revealed viscoelastic residue on the optic body and haptics, and that the lens was received cut in half, which is consistent with a lens that was handled during explant.Based on the return condition of the lens no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
|
| |
|
Additional Manufacturer Narrative
|
|
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are (b)(4) and capa-010215.
|
| |
|