MAUDE Adverse Event Report: HAEMONETICS CORPORATION CELL SAVER ELITE AUTOTRANSFUSION SYSTEM; CS ELITE PROCESSING KIT, 225ML

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

HAEMONETICS CORPORATION CELL SAVER ELITE AUTOTRANSFUSION SYSTEM; CS ELITE PROCESSING KIT, 225ML Back to Search Results
Model Number CSE-P-225
Medical Device Problem Code Fluid/Blood Leak (1250)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/25/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
No patient information was provided at this time.A photo was provided of the bowl from the cell saver elite set which confirmed a leakage at the top header shield ports of the bowl.The bowl is not available for return.Without physical sample provided for evaluation the root cause could not be determined.
 
Event or Problem Description
On (b)(6) 2020 haemonetics was notified of a fluid eruption which had occurred during a spinal procedure, utilizing the cell saver® elite® autotransfusion system and cell saver® elite set - 225ml.There was no reported impact to patients' health and no possible exposure to the techs involved in cleaning the system.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CELL SAVER ELITE AUTOTRANSFUSION SYSTEM
Common Device Name
CS ELITE PROCESSING KIT, 225ML
Manufacturer (Section D)
HAEMONETICS CORPORATION
125 sumer street
boston, ma
MDR Report Key10317739
Report Number1219343-2020-00072
Device Sequence Number515741
Product Code CAC
UDI-Device Identifier10812747016544
UDI-Public(01)10812747016544(17)230401(10)0420022
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K120586
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date (Section B) 06/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/01/2023
Device Model NumberCSE-P-225
Device Lot Number0420022
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/25/2020
Initial Report FDA Received Date07/23/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-