BOSTON SCIENTIFIC CORPORATION VALITUDE X4 CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
|
Back to Search Results |
|
Model Number U128 |
Device Problems
Signal Artifact/Noise (1036); High impedance (1291); Over-Sensing (1438)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 12/10/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
This product has not been returned to boston scientific, and as a result, laboratory analysis could not be conducted.The associated investigation determined that this device exhibited intermittent impedance measurements with no conclusive evidence of a malfunction or inadequate lead-to-device connection; please refer to the description for more information regarding the specific circumstances of this event.
|
|
Event Description
|
It was reported that the right atrial (ra) lead has had a few jumps in pacing impedance measurements, that were greater than 2000 ohms.Later the right ventricular (rv) lead had the same issue.Both leads are from another manufacturer.The leads thus, automatically switched from bipolar to unipolar pacing and sensing.This resulted in myopotential noise on both leads and oversensing only on the ra lead.There was no pacing inhibition.Troubleshooting options were discussed.No adverse patient effects were reported.This product remains in service.This report will be updated, should additional information be received.
|
|