• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS CL; MONOFOCAL IOLS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

JOHNSON & JOHNSON SURGICAL VISION, INC. TECNIS CL; MONOFOCAL IOLS Back to Search Results
Model Number Z9002
Medical Device Problem Codes Difficult to Insert (1316); Defective Component (2292)
Health Effect - Clinical Codes Capsular Bag Tear (2639); Vitrectomy (2643); No Code Available (3191)
Date of Event 05/26/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Age/date of birth: unknown/not provided.Sex/gender: unknown/not provided.If implanted, give date: not applicable, there is no indication the lens was implanted.If explanted, give date: not applicable, as the lens was not implanted; therefore, it was not explanted.(b)(4).Attempts have been made to obtain missing information; however, no definitive response has been received.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that an intraocular lens (iol) had a haptic so big the surgeon could not get it positioned.The capsule tore and a vitrectomy was required, however, it was not product related.It was also reported the patient will be returning aphakic.No additional information was provided.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Additional Manufacturer Narrative
Additional information: section d10: device available for evaluation? yes.Returned to manufacturer on: 08/14/2020.Section h3: device returned to manufacturer ¿ yes.Device evaluation: the z9002 lens was returned in its original package.Visual inspection using magnification was performed to the returned sample: loose fibers/particles were observed on lens relating to handling the unit out of a sterile environment.Residues of lubricant material were also observed on lens.No damaged was observed to the lens.Both haptics were observed damaged/distorted.The sample returned condition is consistent with a lens that was handled and prepared for surgical process.The complaint issue reported could not be verified.Based on the analyzed of the returned sample, there is no indication of a product quality deficiency.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured according to specifications.The search revealed that no similar complaint for this production order number has been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency and the reported issue could not be confirmed.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS CL
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key10318331
Report Number2648035-2020-00548
Device Sequence Number17015318
Product Code HQL
UDI-Device Identifier05050474530768
UDI-Public(01)05050474530768(17)220307
Combination Product (Y/N)Y
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
P880081
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/29/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date03/07/2022
Device Model NumberZ9002
Device Catalogue NumberZ900200235
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2020
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 06/26/2020
Supplement Date Received by Manufacturer08/21/2020
10/25/2020
Initial Report FDA Received Date07/24/2020
Supplement Report FDA Received Date09/16/2020
10/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-