• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 3-4 9MM RT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 3-4 9MM RT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Model Number 74035231
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/26/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that the journey ii bcs articular insert was found to be cracked.No case involved.
 
Additional Manufacturer Narrative
H3, h6: the associated device, intended to use in treatment, was not returned for evaluation.Therefore, product analysis could not be performed at this time.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Based on this investigation, the need for corrective action is not indicated.The device is a reusable instrument that can be exposed to numerous surgeries and cleaning cycles.As plastics are vulnerable and crack may have initiated during use and possible causes could be due to the heating and cooling associated with autoclaving or prolonged use.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II BCS ART ISRT TRL SZ 3-4 9MM RT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10318381
Report Number1020279-2020-03478
Device Sequence Number596573
Product Code JWH
UDI-Device Identifier00885556157480
UDI-Public00885556157480
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 05/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74035231
Device Catalogue Number74035231
Device Lot Number17DAP0024
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/26/2020
Supplement Date Received by Manufacturer05/03/2021
Initial Report FDA Received Date07/24/2020
Supplement Report FDA Received Date05/10/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-