Model Number ZCB00
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code No Code Available (3191)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Date of event - exact date not provided, best estimate is between (b)(6) 2019 - (b)(6) 2020.(b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that a tecnis intraocular lens (iol) was explanted from the patient's operative eye because she is not as nearsighted as she would like during post-op exam.The replacement iol is of the same model with a higher diopter, (20.0).There was no additional surgical intervention performed, and no injury was reported.Visual acuity pre-op, with correction (cc) 20/30 near visual acuity (nva).Visual acuity post-op, sans correction (sc) 20/25 nva.No further information was provided.
 
Additional Manufacturer Narrative
This correction mdr is being submitted to address inaccurate information in section f on the xml version of the previous submission caused by a software glitch in the transmission process.The internal non-conformance numbers for this software issue are nr-0148919 and capa-010215.
 
Additional Manufacturer Narrative
Section d10: device available for evaluation: yes.Section d10: returned to manufacturer on: 7/29/2020.Section h3: device returned to manufacturer: yes.Device evaluation: the product was received inside a non-jnj lens case.Additionally, a complaint intake form was received as well.Visual inspection under magnification revealed viscoelastic residue and what appeared to be dried blood on the optic body and haptics, and that the lens was received cut in half, which is consistent with a lens that was handled during explant.Based on the return condition of the lens no further product evaluation could be performed.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing records review: the manufacturing records for the product were reviewed.The product was manufactured and released according to specification.Historical data analysis: a search of complaints revealed no other complaints have been received for this production order number.Conclusion: as a result of the investigation there is no indication of a quality product deficiency and the reported issue could not be verified.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.H3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS 1-PIECE
Common Device Name
MONOFOCAL IOLS
Manufacturer (Section D)
JOHNSON & JOHNSON SURGICAL VISION, INC.
1700 e st andrew place
santa ana CA 92705
MDR Report Key10318598
Report Number2648035-2020-00546
Device Sequence Number3484771
Product Code HQL
UDI-Device Identifier05050474531666
UDI-Public(01)05050474531666(17)230726
Combination Product (Y/N)Y
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2019
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup,Followup
Report Date (Section B) 10/29/2020
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/26/2023
Device Model NumberZCB00
Device Catalogue NumberZCB0000175
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/29/2020
Was the Report Sent to FDA? No
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 06/26/2020
Supplement Date Received by Manufacturer08/05/2020
10/25/2020
Initial Report FDA Received Date07/24/2020
Supplement Report FDA Received Date08/19/2020
10/29/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Date Report Sent to FDA01/01/2005
-
-