DRÄGERWERK AG & CO. KGAA FABIUS MRI; ANESTHESIA UNITS Back to Search Results
Catalog Number 8607300
Medical Device Problem Codes Gas Output Problem (1266); Loss of Threshold (1633); Failure to Deliver (2338)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/16/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The investigation has just started; results will be provided in a follow-up report.
 
Event or Problem Description
It was reported that the device posted a ventilator failure during use.There was no injury reported.
 
Additional Manufacturer Narrative
The device was inspected on-site by a dräger service engineer who confirmed the reported issue of ventilator failure and replaced the ventilator motor unit upon the findings of initial log file analysis.The device passed all required tests afterwards and could be returned to use.The replaced motor unit was returned to the manufacturer for an in-depth analysis.Visual examination revealed that the piston was blocked at the lower end.Root cause could be traced back to a damaged incremental encoder sensor - as a result from this error condition the software could not recognize the rotation of the motor anymore and the piston was moved towards the lower end position and remained blocked there.This particular failure mode was never observed before in comparison to more than 100.000 devices sold.If automatic ventilation fails the device posts a corresponding alarm to alert the user.Manual ventilation remains possible then.
 
Event or Problem Description
Please refer to initial mfr.Report #9611500-2020-00259.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FABIUS MRI
Common Device Name
ANESTHESIA UNITS
Manufacturer (Section D)
DRÄGERWERK AG & CO. KGAA
moislinger allee 53-55
lübeck 23542
GM  23542
MDR Report Key10319007
Report Number9611500-2020-00259
Device Sequence Number3083996
Product Code BSZ
UDI-Device Identifier04048675041498
UDI-Public(01)04048675041498(11)160909(17)170228(93)8607300-42
Combination Product (Y/N)N
Initial Reporter StateDE
Initial Reporter CountryUS
PMA/510(K) Number
K072884
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
Report Date (Section B) 08/21/2020
Report Date (Section F) 06/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Catalogue Number8607300
Was Device Available for Evaluation? No
Date Report to Manufacturer06/23/2020
Initial Date Received by Manufacturer 06/24/2020
Supplement Date Received by Manufacturer08/12/2020
Initial Report FDA Received Date07/24/2020
Supplement Report FDA Received Date08/21/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-