| Catalog Number |
8607300 |
| Medical Device Problem Codes |
Gas Output Problem (1266); Loss of Threshold (1633); Failure to Deliver (2338)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/16/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation has just started; results will be provided in a follow-up report.
|
| |
|
Event or Problem Description
|
|
It was reported that the device posted a ventilator failure during use.There was no injury reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was inspected on-site by a dräger service engineer who confirmed the reported issue of ventilator failure and replaced the ventilator motor unit upon the findings of initial log file analysis.The device passed all required tests afterwards and could be returned to use.The replaced motor unit was returned to the manufacturer for an in-depth analysis.Visual examination revealed that the piston was blocked at the lower end.Root cause could be traced back to a damaged incremental encoder sensor - as a result from this error condition the software could not recognize the rotation of the motor anymore and the piston was moved towards the lower end position and remained blocked there.This particular failure mode was never observed before in comparison to more than 100.000 devices sold.If automatic ventilation fails the device posts a corresponding alarm to alert the user.Manual ventilation remains possible then.
|
| |
|
Event or Problem Description
|
|
Please refer to initial mfr.Report #9611500-2020-00259.
|
| |
|