• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ACC 2.0MM COCR CABLE W/CLAMP; CERCLAGE, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ACC 2.0MM COCR CABLE W/CLAMP; CERCLAGE, FIXATION Back to Search Results
Model Number 71340007
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 06/28/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery the surgeon used an accord cable, the cable broke during fastening process.There was no injury to patient and a second implant was available.There was no delay reported.
 
Additional Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The cable is broken and some of the wires are fraying, rendering the device inoperative.The device shows signs of extensive use.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACC 2.0MM COCR CABLE W/CLAMP
Common Device Name
CERCLAGE, FIXATION
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10319274
Report Number1020279-2020-03483
Device Sequence Number18981686
Product Code JDQ
UDI-Device Identifier03596010485731
UDI-Public03596010485731
Combination Product (Y/N)N
Initial Reporter CountrySF
PMA/510(K) Number
K031162
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71340007
Device Catalogue Number71340007
Device Lot Number18GSN0490
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2021
Initial Date Received by Manufacturer 06/29/2020
Supplement Date Received by Manufacturer10/21/2021
Initial Report FDA Received Date07/24/2020
Supplement Report FDA Received Date10/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-