| Model Number |
MMT-723NAB |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); Moisture or Humidity Problem (2986)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/18/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump got wet and now it is not working.Customer jumped in the water and forgot to take it off and it gave a stuck button error.The customer stated was unable to clear the alarm.No harm requiring medical intervention was reported.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with unresponsive buttons due to moisture damage at keypad traces.No button error alarm noted.Traces of corrosion found at the electronic assembly and motor assembly per visual inspection.Unable to perform the displacement and self test due to unresponsive buttons.However, keypad flattened button dome switch (creased) was found on esc and act.
|
| |
|