MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR PAIN RELIEF
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
Overheating of Device (1437); Charging Problem (2892)
|
Patient Problem
Burn, Thermal (2530)
|
Event Date 07/19/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 97755, serial#: (b)(4), product type: recharger.Other relevant device(s) are: product id: 97755, serial/lot #: (b)(4), udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a consumer regarding a patient with an implantable neurostimulator (ins).It was reported the patient was having problems with their equipment since sunday, prior to (b)(6) 2020.The patient stated the recharger box was getting too hot and it burned when it touched their skin.The patient received no device found when charging the ins.A new recharger was requested.No further complications were reported or anticipated.
|
|
Manufacturer Narrative
|
H3: the recharger, s/n (b)(6) was returned for product analysis.Analysis of the device found that the recharge antenna insulation was pulled away, cable insulation is peeling away at donut end and wires are exposed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|