• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY MITEK LLC US 5.5 HEALIX ADVANCE KNTLSS PK; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY MITEK LLC US 5.5 HEALIX ADVANCE KNTLSS PK; SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE Back to Search Results
Model Number 222334
Medical Device Problem Codes Mechanical Jam (2983); Migration (4003)
Health Effect - Clinical Code Not Applicable (3189)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Udi: (b)(4).As of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported by the sales rep that during a rotator cuff repair surgery on an unknown date, it was observed that the 5.5 healix advance kntlss pk device became stuck in the inserter after implanting for a lateral row.Subsequently, the anchor came out while attempting to pull the inserter out.There was a surgical delay of two minutes to complete the surgery with another like device.There were no fragments generated.There were no patient consequences reported.No additional information was provided.
 
Additional Manufacturer Narrative
This report is being submitted in pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by mitek or its employees that the report constitutes an admission that the device, mitek, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: correction: g1-2: the manufacturing location has been updated to reflect the correct information.Additional information: investigation summary
=
> the complaint device is not being returned, it was discarded by the customer, therefore unavailable for a physical evaluation.With the information provided, and without the complaint device to evaluate, we cannot determine a root cause for the reported failure.A manufacturing record evaluation was performed for the finished device lot number (l991807), and no non-conformances were identified.At this point in time, no corrective action is required, and no further action is warranted.However, depuy synthes mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
5.5 HEALIX ADVANCE KNTLSS PK
Common Device Name
SOFT-TISSUE ANCHOR, NON-BIOABSORBABLE
Manufacturer (Section D)
DEPUY MITEK LLC US
325 paramount drive
raynham MA 02767
MDR Report Key10319982
Report Number1221934-2020-01903
Device Sequence Number14219374
Product Code MBI
UDI-Device Identifier10886705023486
UDI-Public10886705023486
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K130539
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/31/2021
Device Model Number222334
Device Catalogue Number222334
Device Lot NumberL991807
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/09/2020
Supplement Date Received by Manufacturer07/30/2020
Initial Report FDA Received Date07/24/2020
Supplement Report FDA Received Date07/30/2020
Patient Sequence Number1
-
-