| Model Number |
CD3257-40 |
| Medical Device Problem Codes |
Failure to Interrogate (1332); Inappropriate/Inadequate Shock/Stimulation (1574); Inappropriate or Unexpected Reset (2959)
|
| Health Effect - Clinical Code |
Therapeutic Effects, Unexpected (2099)
|
| Date of Event |
07/12/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
Related manufacturer reference number: 2938836-2020-07618.It was reported that the patient presented in emergency room, inappropriate shocks due to backup vvi issue was observed.The device was explanted and replaced.The patient was stable.
|
| |
|
Event or Problem Description
|
|
New information received notes that failure to interrogate was observed.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: d10, h6.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported field event of backup vvi (bvvi) was confirmed in the laboratory due to event queue overflow caused by an rf ic anomaly.The reported event of inappropriate shock and failure to interrogate could not be confirmed.The device was in backup mode upon receipt, so a full normal interrogation could not be performed.However, the device¿s telemetry was normal.The device was able to be interrogated via both inductive and rf telemetry during testing in the laboratory.
|
| |
|