| Model Number |
FRED3021-PMA |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Codes |
No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/26/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with this part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The delivery pusher was returned for evaluation.Upon initial investigation of the pusher, it did not appear to be straight.The reported complaint is non-verifiable as the stent wasn't returned to determine if it would get caught.The root cause of this complaint may have been due to tortuous anatomy causing the pusher to become stuck on the stent during removal.It may have also resulted in the pusher not being straight due to the tortuosity.The investigation of the returned pusher found no condition that would have directly resulted in the reported event.
|
| |
|
Event or Problem Description
|
|
It was reported that treatment was performed for an anterior communicating artery (acom) aneurysm.The origin of the acom complex was reported to be slightly tortuous with some narrowing near the neck of the aneurysm.After placement of the fred between the a2 and a1 segment, the pusher wire caught on the stent during the removal attempt.The intermediate catheter was advanced and used to detach the stent from the pusher wire and microcatheter.After deployment, the fred slightly extended into the internal carotid artery.The stent remains implanted.No additional intervention was performed and there was no patient injury.
|
| |
|
Additional Manufacturer Narrative
|
|
The stent was not returned for analysis, as it remains implanted.Investigation of the returned delivery pusher revealed the distal marker band was scratched, which may have occurred during retrieval of the pusher from the stent.Fluoroscopic images demonstrate the pusher positioned at approximately 90 degrees to the stent, which may have caused the distal marker band to get caught onto the device; however, without examination of the stent, it cannot be determined if it caused or contributed to the alleged product issue.
|
| |
|
Search Alerts/Recalls
|