| Model Number |
72200617S |
| Medical Device Problem Code |
Mechanical Jam (2983)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
| Date of Event |
05/15/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The analysis performed by the manufacturer has not concluded yet.
|
| |
|
Event or Problem Description
|
|
It was reported the motor drive unit was not working.Incident occurred before the procedure; therefore, there was no patient involvement.A preliminary analysis performed by the manufacturer revealed a blade stall error, which is a reportable event.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: h2, h3, h6.The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A visual inspection was performed on the product and observed no issues.A functional evaluation revealed a blade stall error.The complaint was confirmed and the root cause has been associated with a mechanical component failure.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found related failures; this failure mode will be trended to assess for any necessary corrective actions.Factors that could have contributed to the reported event include corrosion in the gearbox/motor assembly from cleaning and sterilization methods and the chemicals involved over a period of time or one or more of the motor phases shorting out.
|
| |
|