Model Number 201-90411
Device Problems Overheating of Device (1437); Decreased Pump Speed (1500); Communication or Transmission Problem (2896)
Patient Problem No Patient Involvement (2645)
Event Date 06/29/2020
Event Type  malfunction  
Manufacturer Narrative
No further information was provided.A supplemental report will be submitted when the manufacturer's investigation is completed.
 
Event Description
Related manufacturer report number: 2916596-2020-03409.It was reported that the centrimag motor could not be recognized by the system.The site communicated that the motor could not operate above 4500 rotations per minute and warm to the touch.
 
Manufacturer Narrative
Section d4: additional information section g9: the initial report was sent with an incorrect first mfr number (2916596).The correct number is (b)(6).Manufacturer's investigation conclusion: the reported event of the motor not being recognized by the system and being warm to the touch was not confirmed.The centrimag 2nd generation console was not returned for analysis.Additional provided information communicated on (b)(6)2020 stated that the serial number of the console was not known.The console worked fine when connected to another motor.The root cause for the reported event was unable to be conclusively determined through this analysis.The 2nd generation centrimag system operating manual section 4 entitled "warnings & precautions" warns "one additional 2nd generation centrimag primary console, motor and flow probe are required as backup system in the immediate vicinity of each patient whenever the centrimag or pedivas blood pump is used.The backup console must be connected to the backup motor and to the backup flow probe, have a battery charge sufficient for at least one hour of operation, be connected to ac power (except during transport) and be immediately available should the main console, motor or flow probe experience a malfunction." the 2nd generation centrimag system operating manual section 10 entitled "emergency and troubleshooting" states that "the recommended practice whenever there is a 2nd generation centrimag primary console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the backup motor and console to continue patient support.Do not exchange individual motors or individual consoles during patient support." the 2nd generation centrimag system operating manual has an emergency/troubleshooting section for the 2nd generation console.The recommended practice whenever there is a console or motor malfunction is to replace the console and motor as a set.Remove the blood pump from the malfunctioning motor and console and place the blood pump in the back-up motor and console.Switch all components (console, motor, flow probe and cables) simultaneously to continue patient support, and then perform troubleshooting on the non-functioning system, when it is no longer being used for patient support.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CENTRIMAG 2ND GENERATION PRIMARY CONSOLE
Type of Device
CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
THORATEC SWITZERLAND GMBH
technoparkstrasse 1
zurich CH-80 05
SZ  CH-8005
MDR Report Key10321472
MDR Text Key202268783
Report Number2916596-2020-03772
Device Sequence Number1
Product Code DWA
Combination Product (y/n)N
PMA/PMN Number
K131179
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number201-90411
Device Catalogue Number201-90411
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received10/21/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
CENTRIMAG MOTOR, US.
-
-