• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 7-8 9MM LT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/META

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. JRNY II BCS ART ISRT TRL SZ 7-8 9MM LT; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/META Back to Search Results
Model Number 74035281
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/30/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that upon inspection the device was found with gouges/cracks/scratches.No case involved.
 
Additional Manufacturer Narrative
The device, intended use in treatment, was returned for evaluation.A visual inspection confirmed the jrny ii bcs art isrt trl sz 7-8 9mm lt shows nicks and scratches.The device shows significant signs of wear/usage.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we have no reason to suspect that the products failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.This is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation warranted for this complaint; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JRNY II BCS ART ISRT TRL SZ 7-8 9MM LT
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/META
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10322136
Report Number1020279-2020-03509
Device Sequence Number2678663
Product Code JWH
UDI-Device Identifier00885556157336
UDI-Public00885556157336
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
K111711
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number74035281
Device Catalogue Number74035281
Device Lot Number13KAP6183
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/20/2020
Initial Date Received by Manufacturer 06/30/2020
Supplement Date Received by Manufacturer07/31/2020
Initial Report FDA Received Date07/24/2020
Supplement Report FDA Received Date08/04/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-