• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COOK INC UNKNOWN; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COOK INC UNKNOWN; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number UNKNOWN
Device Problems Structural Problem (2506); Device Tipped Over (2589); Unintended Movement (3026); Insufficient Information (3190)
Patient Problems Internal Organ Perforation (1987); Pain (1994); Numbness (2415); No Information (3190)
Event Type  Injury  
Manufacturer Narrative
Occupation: non-healthcare professional.Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog number and lot number are unknown, however, the alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
 
Event Description
It is alleged that the patient received a gunther tulip on (b)(6) 2005.It is alleged that the patient was injured without further explanation.Hospital and medical records have been requested, but not yet provided.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.The following fields were updated per additional information received: a2, b1, b2, b5, b6, and h6.Investigation.The following allegations have been investigated: organ perforation, migration, tilt, leg pain, feet numbness.Investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Filter or filter fragment migration and (or) embolization (e.G., movement to the heart or lungs) has been reported.Filter or filter fragment movement has occurred in both the cranial and caudal direction and may be either symptomatic or asymptomatic.Potential causes may include filter placement in ivcs with diameters smaller or larger than those specified in these instructions for use; improper deployment; deployment into thrombus; dislodgement due to large thrombus burdens; and (or) excessive force or manipulations near an in situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: filter migration, trauma to adjacent structures.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported leg pain and feet numbness are directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog number and lot number are unknown; however, the alleged tulip is manufactured and inspected according to controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.This report is required by the fda under 21 cfr part 803.This report is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement made in it is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, or that any cook device caused or contributed to or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Patient allegedly received an implant on (b)(6) 2005 via the right common femoral vein due to an accident.The patient alleges organ perforation.The patient further alleges leg pain, numbness in feet.(b)(6) 2019, per a report from computed tomography; ¿the ivc filter is tilted 17 [degrees] medially with the apex migrated to the level of the left renal vein and 1cm above the right renal vein.The apex abuts the ivc wall.Anterior strut extends 2mm into the duodenal wall.The medial, lateral and posterior struts extending 2mm into the mesenteric fat but do not invade any organ.Fracture or bending of the struts.No ivc stenosis.¿.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
COOK INC
750 daniels way
bloomington IN 47404
MDR Report Key10322925
MDR Text Key200967330
Report Number1820334-2020-01362
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Type of Report Initial,Followup
Report Date 10/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/20/2020
Initial Date FDA Received07/24/2020
Supplement Dates Manufacturer Received09/24/2020
Supplement Dates FDA Received10/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Life Threatening; Other;
-
-