Model Number 466FXXXX
Device Problem Difficult to Remove (1528)
Patient Problem Perforation of Vessels (2135)
Event Date 06/03/2020
Event Type  Injury  
Manufacturer Narrative
As reported, the patient underwent placement of the optease inferior vena cava (ivc) filter.The indication for filter placement is not available.The filter subsequently malfunctioned and caused injury and damages to the patient, including, but not limited to, physical and emotional damages from the filter perforating the wall of the ivc and a failed retrieval attempt.The filter remains implanted; thus, unavailable for analysis.The product was not returned for analysis and the sterile lot number has not been provided; therefore, no device analysis nor device history record review could be performed.The optease vena cava filter is indicated for use in the prevention of recurrent pulmonary embolism (pe) via percutaneous placement in the vena cava for patients in which anticoagulants are contraindicated, anticoagulant therapy for thromboembolic disease has failed, emergency treatment following massive pulmonary embolism where anticipated benefits of conventional therapy are reduced or for chronic, recurrent pulmonary embolism where anticoagulant therapy has failed, or is contraindicated.The purpose of a vena cava filter is to catch thrombus from the lower extremities as it travels along normal blood flow patterns up towards the heart.Without procedural films for review, the reported embedded in ivc wall and retrieval difficulty could not be confirmed and the exact cause could not be determined.Retrieval of the optease vena cava filter is indicated, in the us, up to 14 days post implantation.Usage of the product other than that indicated in the product's ifu may involve additional risks not described in the labeling.The predominant concern is the development of endothelialization, which would make subsequent removal difficult.Endothelialization has been shown to lead to explantation problems after as short a period as 12 days.It was reported that there was perforation of the ivc; however, a clinical conclusion could not be determined as to the cause of the event.A review of the instructions for use notes vessel damage such as intimal tears and perforation as procedural complications related it ivc filters.Ivc perforation from removable filters is relatively common, and directly related to how long the filter has been in place.Studies have noted a greater than 80% perforation rate overall, with all filters imaged after 71 days from implantation revealing some level of perforation.Clinical factors that may have influenced the event include patient, pharmacological and lesion characteristics.Without procedural films or images for review the reported event(s) could not be confirmed.Given the limited information available for review at this time, there is nothing to suggest that the reported events are related to the design and manufacturing process of the device; therefore, no corrective action will be taken.Should additional information become available, the file will be updated accordingly.
 
Event Description
As reported by the legal team, the patient underwent placement of the optease vena cava filter.The filter subsequently malfunctioned and caused injury and damages to the patient, including, but not limited to, physical and emotional damages from the filter perforating the wall of the ivc and a failed retrieval attempt.As a direct and proximate result of these malfunctions, the patient suffered life-threatening injuries and damages, and required extensive medical care and treatment.As a further proximate result, the patient has suffered and will continue to suffer significant medical expenses, and pain, suffering, loss of enjoyment, distress and other damages.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
466FXXXX
Type of Device
FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
CORDIS CASHEL
cahir road
cashel, tipperary 0000
EI  0000
Manufacturer (Section G)
CORDIS CASHEL
cahir road
cashel, tipperary 0000
EI   0000
Manufacturer Contact
karla castro
14201 nw 60th ave
miami lakes, FL 33014
7863138372
MDR Report Key10323351
MDR Text Key200427553
Report Number1016427-2020-04229
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K020316
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 07/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number466FXXXX
Device Catalogue Number466FXXXX
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/08/2020
Initial Date FDA Received07/24/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
-
-