DEPUY ORTHOPAEDICS INC US DUR MAR +4 NEUT LNR 36X56; DURALOC IMPLANT : HIP POLY ACETABULAR LINERS Back to Search Results
Model Number 1220-36-456
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Hypersensitivity/Allergic reaction (1907); No Code Available (3191)
Date of Event 07/13/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient had pseudo tumor and corrosion at the head neck taper joint.Revision was performed due to presence of said pseudo tumor and corrosion with elevated metal ion levels.A head and liner swap was performed.Stem and cup were originally implanted on (b)(6) 1997.The first revision was done on 2015 due to poly wear and this was the second revision.No implant records for the stem, cup or head.Doi: (b)(6) 1997 - stem and cup, doi: (b)(6) 2015 - head and liner, dor: (b)(6) 2020, affected side: right hip.
 
Additional Manufacturer Narrative
Product complaint # : (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint (b)(4).Investigation summary :no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DUR MAR +4 NEUT LNR 36X56
Common Device Name
DURALOC IMPLANT : HIP POLY ACETABULAR LINERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10323769
Report Number1818910-2020-16873
Device Sequence Number2951259
Product Code LPH
UDI-Device Identifier10603295015703
UDI-Public10603295015703
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
K010171
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2015
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 07/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1220-36-456
Device Catalogue Number122036456
Device Lot Number591914
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/13/2020
Supplement Date Received by Manufacturer07/13/2020
08/24/2020
Initial Report FDA Received Date07/24/2020
Supplement Report FDA Received Date07/30/2020
08/25/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ARTICULEZE M HEAD 36MM +8.5; DUR DYNAMIC LOCK RING 56 OR 68; DUR MAR +4 NEUT LNR 36X56; UNKNOWN HIP ACETABULAR CUP; UNKNOWN HIP FEMORAL STEM
Outcome Attributed to Adverse Event Required Intervention;
Patient Age67 YR
Patient Weight65
-
-