Model Number 15 |
Device Problems
Failure to Deliver Shock/Stimulation (1133); Inappropriate/Inadequate Shock/Stimulation (1574)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 06/30/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Physio-control evaluated the customer¿s device and verified the reported issue.Physio replaced the therapy pcb assembly to resolve this issue.After other unrelated repairs were completed, proper device operation was observed through functional and performance testing and the device was returned to the customer for use.
|
|
Event Description
|
The customer contacted physio-control to report that their device failed the user test.Upon evaluation of the customer's device, physio-control observed that an event code logged in the device's memory that is related to a potential failure of the device to deliver defibrillation energy.Physio also observed another event code logged in the device's memory that is indicative of a device failure that could result in a partial loss of defibrillator output energy due to a loss of the negative portion of the biphasic output waveform, resulting in a monophasic shock.Additionally, the defibrillator output energy may be reduced by up to approximately 20% from the selected energy level.As a result, the wrong defibrillation therapy would be delivered, if it were necessary.There was no patient use associated with the reported event.
|
|
Manufacturer Narrative
|
Physio-control further evaluated the removed therapy pcb assembly and determined the cause of the reported issue was due to transistor, designator q7.Q7 had become electrically shorted.
|
|
Event Description
|
The customer contacted physio-control to report that their device failed the user test.Upon evaluation of the customer's device, physio-control observed that an event code logged in the device's memory that is related to a potential failure of the device to deliver defibrillation energy.Physio also observed another event code logged in the device's memory that is indicative of a device failure that could result in a partial loss of defibrillator output energy due to a loss of the negative portion of the biphasic output waveform, resulting in a monophasic shock.Additionally, the defibrillator output energy may be reduced by up to approximately 20% from the selected energy level.As a result, the wrong defibrillation therapy would be delivered, if it were necessary.There was no patient use associated with the reported event.
|
|
Search Alerts/Recalls
|