| Model Number |
MMT-723NAS |
| Medical Device Problem Codes |
Device Difficult to Program or Calibrate (1496); Power Problem (3010)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that insulin pump had buttons were hard to pressed.No harm requiring medical intervention was reported.The insulin pump had to reprogrammed time after battery change and rejected new battery.The device will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with intermittent button response due to partial unlocked lcd keypad connector noted during visually inspect.Unable to perform all functional testing including the displacement, operating currents, a21 error test, rewind, basic occlusion, occlusion, prime or a33 and excessive no delivery test or verify failed battery test alert due to buttons not responding.(b)(4).
|
| |
|