MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REAMER DOME 50MM; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REAMER DOME 50MM; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71362750
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Known Impact Or Consequence To Patient (2692)
Date of Event 06/29/2020
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that during surgery a r3 50mm reamer dome broke outside of the patient after being used.A s&n back up was available.The case was not delayed and there was not an adverse event with regards to the patient.
 
Additional Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection of the returned reamer dome confirms the cross bar broke off the device.All broken pieces were returned for evaluation.The device was manufactured in 2009.The device exhibits signs of significant wear and use.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REAMER DOME 50MM
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10324347
Report Number1020279-2020-03520
Device Sequence Number14122878
Product Code JDG
UDI-Device Identifier03596010456885
UDI-Public03596010456885
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/10/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71362750
Device Catalogue Number71362750
Device Lot Number09CM19564
Initial Date Received by Manufacturer 06/29/2020
Supplement Date Received by Manufacturer08/06/2020
Initial Report FDA Received Date07/25/2020
Supplement Report FDA Received Date08/10/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-