| Model Number |
MMT-523LNAL |
| Medical Device Problem Codes |
Failure to Cycle (1142); Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/20/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that insulin pump had compromised force sensor system alarm.No harm requiring medical intervention was reported.The customer was assisted with troubleshooting.Customer stated drive support cap was normal and pressing the cap while did not connected.The device will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device received with the drive support disk was inspected and found loose or protruded and broken reservoir tube lip.A test reservoir was installed and did not lock in place due to complete broken reservoir tube lip.Unable to perform all functional testing including the displacement, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test or verify a33 alarm due to completely broken reservoir tube lip.
|
| |
|