| Model Number |
MMT-523NAS |
| Medical Device Problem Codes |
Mechanical Problem (1384); Battery Problem (2885)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/21/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had motor error alarm.The customer stated that insulin pump rejected new batteries.No harm requiring medical intervention was reported.The insulin pump will not be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed all functional testing including the displacement, rewind, basic occlusion, occlusion, prime/a33 and excessive no delivery test.No unexpected low battery alarm anomalies noted.No excessive no delivery/occlusion alarm noted.No motor error alarm noted.Motor passed motor test.(b)(4).
|
| |
|