| Model Number |
ME2017 |
| Medical Device Problem Code |
Insufficient Device Problem Information (3190)
|
| Health Effect - Clinical Code |
No Information (3190)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
An unexpected return confirmed a male luer break.The set was visually inspected for cracks, misassembly or damages to the components.Visual inspection of the set noted that a male luer collar and tip of the extension set was cracked and broken.Examination under magnification observed no stress marks at the break site of the male luer.No tool marks were observed.Functional testing was deemed unnecessary due to the broken male luer collar.The root cause was identified as related to design of the specific male luer component.Device history record for model me2017 could not be performed due to no lot number being provided by customer.
|
| |
|
Event or Problem Description
|
|
Although an unexpected tubing set with an unspecified issue was returned by the customer, a photo from the receiving lab appeared to show a broken luer.No further information is available.
|
| |
|