| Catalog Number |
14-592540 |
| Medical Device Problem Codes |
Fracture (1260); Noise, Audible (3273)
|
| Health Effect - Clinical Codes |
Pain (1994); Inadequate Osseointegration (2646); Foreign Body In Patient (2687); No Code Available (3191)
|
| Date of Event |
01/22/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for evaluation but photos of the device were provided.The screw appears fractured in a manner consistent with a fatigue fracture.The complaint is confirmed.A review of the manufacturing records did not identify any issues which would have contributed to this event.The labeling was reviewed and found to contain instructions regarding proper device usage.Reference report 3012447612-2020-00313.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent a revision surgery after two polaris screws were found to have fractured postoperatively at s1.The patient reported increasing back pain, left leg radiculopathy and a creaking sound from the implants prior to the revision.The fractured portion of the screw was removed but the shaft was retained by the patient.The level was reinstrumented after the surgeon found no fusion present.The patient reports ongoing back pain after the revision.This is report two of two for this event.
|
| |
|