A supplemental report is being submitted for device evaluation product event summary: the ventricular assist device (vad) was returned for evaluation.Various analyses were conducted and reviewed in order to evaluate the performance of the device in relation to the reported event.A review of the manufacturing documentation confirmed that the associated device met all requirements for release.Information provided by the site revealed that the patient was experiencing low flows; the pump speed was adjusted and the patient was started on milrinone.It was suspected that the ventricular assist device (vad) was mispositioned and the vad was explanted.The reported low flow event was not confirmed via review of the controller log files since log files covering the reported event date were not available for analysis.The reported pump malposition event could not be confirmed due to insufficient evidence.Failure analysis of the returned pump revealed that the device passed functional testing and dimensional verification.Visual inspection revealed damage at the connection between the strain relief and header.This is an additional observation not related to the reported event.Given that this damage was not present prior to release of the device, it can like ly be attributed to the handling of the device during or after the explant procedure.Internal pathological report revealed no evidence of thrombus within the device.There is no evidence of a malfunction that may have prevented the pump from performing as intended.Based on risk documentation, multiple factors may have contributed to the low flow event including but not limited to thrombus at the inflow cannula/outflow graft, constriction at the outflow graft, poor vad filling, and/or inappropriate pump rotational speed.Possible clinical factors that may have contributed to this event include the patient¿s pre-existing history and related comorbidities, the progression of their underlying disease, issues related to the therapeutic use of anticoagulant and antiplatelet medications and the patient's complex post-operative course.There are possible patient, pharmacological and procedural factors that may have contributed to this event.Investigation of this event is completed, and the file will be closed.If new information is received, the file will be re-opened, and a supplemental will be submitted.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
A supplemental is being submitted for a correction.Corrected field: h6 fda patient code was updated to e2402.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|