| Model Number |
5400201000 |
| Medical Device Problem Code |
Unintended Power Up (1162)
|
| Health Effect - Clinical Codes |
No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 3 events were reported for this quarter.Product return status: 2 devices were received.1 device investigation type has not yet been determined.Event confirmation status: 2 reported events were confirmed.Evaluation results: 2 devices were found to be affected by a loose lever assembly.3 devices were not labeled for single-use.3 devices were not reprocessed or reused.
|
| |
|
Event or Problem Description
|
|
This report summarizes 3 malfunction events in which the device ran without user activation.3 events had no patient involvement; no patient impact.
|
| |
|
Additional Manufacturer Narrative
|
|
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.3 previously reported events are included in this follow-up record.Product return status 3 devices were received.
|
| |
|
Event or Problem Description
|
|
This report summarizes 3 malfunction events in which the device ran without user activation.- 3 events had no patient involvement; no patient impact.
|
| |
|
Search Alerts/Recalls
|