STRYKER INSTRUMENTS-KALAMAZOO SYSTEM 7 V-NOTCH SAGITTAL SAW; INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT Back to Search Results
Model Number 7208008000
Medical Device Problem Code Fail-Safe Problem (2936)
Health Effect - Clinical Code No Patient Involvement (2645)
Date of Event 06/30/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This record is a consolidation of records summarized as a part of the fda voluntary malfunction summary reporting program.Reported events: 1 event was reported for this quarter.Product return status: 1 device was received.Event confirmation status: 1 reported event was confirmed.Evaluation results: 1 device was found to be affected by a contaminated trigger.1 device was not labeled for single-use.1 device was not reprocessed or reused.
 
Event or Problem Description
This report summarizes 1 malfunction event in which the device experienced a fail-safe problem that could result in unintended activation.1 event had no patient involvement; no patient impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SYSTEM 7 V-NOTCH SAGITTAL SAW
Common Device Name
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED MOTOR AND ACCESSORY/ATTACHMENT
Manufacturer (Section D)
STRYKER INSTRUMENTS-KALAMAZOO
4100 east milham avenue
kalamazoo MI 49001
MDR Report Key10330252
Report Number0001811755-2020-01924
Device Sequence Number11656179
Product Code HWE
UDI-Device Identifier07613252251040
UDI-Public07613252251040
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)Y
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 07/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number7208008000
Device Catalogue Number7208008000
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 06/30/2020
Initial Report FDA Received Date07/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-