Model Number 3660 |
Device Problem
Premature Elective Replacement Indicator (1483)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/09/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
He device is included in the neuromodulation implantable pulse generator (ipg) inaccurate elective replacement indicator advisory notice issued by abbott on 12 september 2017.The results/ method and conclusion codes along with investigation results will be provided in the final report.
|
|
Event Description
|
It was reported the external device displayed early replacement indicator (eri) message.Wireless software update was performed clearing the eri message.The device is providing therapy.
|
|
Manufacturer Narrative
|
An elective replacement indicator message was reported to abbott.The device was not returned for analysis; however, event details indicate that the system was updated to resolve the issue.Actions have been taken to prevent reoccurrence.
|
|