• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. OER-4 100V; ENDOSCOPE REPROCESSOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. OER-4 100V; ENDOSCOPE REPROCESSOR Back to Search Results
Model Number OER-4
Device Problem Device Reprocessing Problem (1091)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/01/2020
Event Type  malfunction  
Manufacturer Narrative
The subject device was not returned to olympus medical systems corp.(omsc), therefore omsc cannot investigate the subject device.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.When omsc checked the use log of the subject device, there was no error about the concentration of the disinfectant solution.The exact cause of the reported event could not be conclusively determined.If additional information becomes available, this report will be supplemented.
 
Event Description
The user alleged that disinfectant solution used for this device got expired too soon.The user tested the concentration level of the disinfectant solution before running the device's reprocessing cycle and confirmed that it had an effective value.After three reprocessing cycles, the user tested the concentration level again and the test results indicated it was expired.The customer was concerned about a possibility that scopes might have not been reprocessed properly.There was no patient injury report related to the event, including infections.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OER-4 100V
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10330551
MDR Text Key223253848
Report Number8010047-2020-04871
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 07/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberOER-4
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/01/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/18/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-