| Model Number |
BD710FJ282CT |
| Medical Device Problem Code |
Sharp Edges (4013)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
06/30/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation details: the device evaluation has been completed.The device was visually inspected and it was found in good conditions.The electrodes were inspected and are in good conditions, near the electrode #1 has the anchor slot that is part of the design and is in normal conditions.A manufacturing record evaluation was performed and no internal actions related to the reported complaint were identified.The customer complaint cannot be confirmed.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer¿s ref # (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that a patient underwent atrial flutter right (r-afl) ablation procedure with a webster ® cs catheter with ez steer® technology and auto id where damaged electrode issue occurred.It was reported that the distal pole had a partial electrode.When removing out of the body looks like the electrode is missing or may be buried in the plastic.The cable was replaced without resolution.A different port was also tried but the issue persisted.The webster ® cs catheter with ez steer® technology and auto id was replaced and the issue resolved.The case continued.There was no report of patient consequence or extended hospitalization.
|
| |
|