Model Number MMT-523LNAS
Medical Device Problem Codes Display or Visual Feedback Problem (1184); Device Difficult to Program or Calibrate (1496); Power Problem (3010); Patient Device Interaction Problem (4001)
Health Effect - Clinical Code No Consequences Or Impact To Patient (2199)
Date of Event 07/22/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.
 
Event or Problem Description
Information received by medtronic indicated that the act button was not responding.Customer stated that the insulin pump had to rewind more than once.Customer stated that the screen was scratched.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
 
Additional Manufacturer Narrative
Device received with all buttons responding properly.No damage on keypad assembly.No unlocked lcd keypad connector.Device received with all operating currents within specification and passed a21 error test and displacement test.No unexpected battery out limit alarm anomalies noted.No unexpected back light anomalies noted.Device received with minor scratched lcd window and cracked battery tube threads.The test infusion set and reservoir does lock into place.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARADIGM REAL-TIME REVEL INSULIN INFUSION PUMP MMT-523LNAS
Common Device Name
PUMP, INFUSION, INSULIN, TO BE USED WITH INVASIVE GLUCOSE SENSOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
MDR Report Key10331532
Report Number2032227-2020-153557
Device Sequence Number3574465
Product Code OYC
UDI-Device Identifier00643169513860
UDI-Public(01)00643169513860
Combination Product (Y/N)N
Initial Reporter StateOH
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 10/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Lay User/Patient
Device Model NumberMMT-523LNAS
Device Catalogue NumberMMT-523LNAS
Device Lot NumberA4523LNASJ
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/15/2020
Initial Date Received by Manufacturer 07/22/2020
Supplement Date Received by Manufacturer10/26/2020
Initial Report FDA Received Date07/28/2020
Supplement Report FDA Received Date10/27/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age71 YR
Patient Weight77
-
-