| Model Number |
R SERIES |
| Medical Device Problem Code |
No Display/Image (1183)
|
| Health Effect - Clinical Code |
No Patient Involvement (2645)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
| |
|
Event or Problem Description
|
|
Complainant alleged that during biomed testing, the device powered on without display.Complainant indicated that there was no patient involvement in the reported malfunction.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned to zoll medical corporation.The customer's report was verified and attributed to a faulty integrated circuit on the digital board.The digital board was replaced to remedy the report.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
| |
|