| Model Number |
I.V. STATION |
| Medical Device Problem Codes |
Incorrect Measurement (1383); Incorrect, Inadequate or Imprecise Result or Readings (1535); Output Problem (3005)
|
| Health Effect - Clinical Code |
No Known Impact Or Consequence To Patient (2692)
|
| Date of Event |
06/29/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
On (b)(6) 2020 an i.V.Station device passed three underdosed syringe preparations.The syringes contained approximately 4ml of neostigmine drug solution when they should have contained 5ml.No errors were generated and the incorrect volume issue was not able to be reproduced.A system modification was made to enable the robot pincer pressure regulator in order to ensure that the syringe is not handled with excessive pressure causing potential drug loss after the preparation has been weighed and marked acceptable.The preparations were visually identified and rejected and there are no known adverse patient effects.
|
| |
|
Additional Manufacturer Narrative
|
|
As of january 2020, the establishment registration and listing for this device was updated to omnicell, inc.From aesynt, inc, which was not reflected in the original report submission.Therefore, this report is a correction to the manufacturer listed in sections d3 and g1.
|
| |
|