Initial reporter is a siemens affiliate.Reporting facility telephone number is: (b)(6).Siemens investigated the reported event.This problem is related to the following software/workflow: syngo.Via rt image suite can be used as advanced application on somaris10 ct scanners.It offers the feature to do patient marking with an external laserbridge in conjunction with a ct scan for cancer treatment planning.To determine the correct mechanical distance between the laserbridge and the isocenter of the ct scanner gantry, the user must perform a new calibration and to record the measured "calibration offset parameter".All further actions will be reported to the fda separately as a voluntary recall action, starting with the customer safety advisory notice (csan) sy051/20/s.
|
|
It was reported to siemens that due to a software problem, when the software version is updated from (b)(4), the "calibration offset parameter" which determines the distance between the external lasers for patient marking and the ct central image plane, is silently overwritten to the factory default settings (50.0cm).As of the date of this report, no injury has been reported to siemens.In a worst-case scenario, if the user doesn't realize this change and does not perform a re-calibration, this change may result in an incorrect patient marking and subsequently to a radiation dose applied to wrong location during the cancer treatment.This worst-case scenario can only occur if some other process steps such as, position verification at the linear accelerator before the treatment, do not clearly show any deviation.This problem is related to the following software/workflow: syngo.Via rt image suite can be used as advanced application on somaris10 ct scanners.The reported event occurred in (b)(6).
|