| Model Number |
IPN000188 |
| Medical Device Problem Code |
Difficult to Remove (1528)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/09/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that after expanding and deflating the balloon, the balloon was stuck inside the arrow 8f short sheath.As a result, the sheath with the stuck balloon was removed as a whole, leaving the guide wire in and inserting another sheath and a new balloon was used.There was no report of patient complications, serious injury or death.
|
| |
|
Additional Manufacturer Narrative
|
|
(b)(4).No part has returned to teleflex chelmsford for investigation.The reported complaint of catheter tight in sheath is not able to be confirmed.If the product is returned at a later date, a full investigation of the sample will be completed.The root cause of the complaint is undetermined.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revived risk.This will be monitored for any developing trends.
|
| |
|
Event or Problem Description
|
|
It was reported that after expanding and deflating the balloon, the balloon was stuck inside the arrow 8f short sheath.As a result, the sheath with the stuck balloon was removed as a whole, leaving the guide wire in and inserting another sheath and a new balloon was used.There was no report of patient complications, serious injury or death.
|
| |
|