MAUDE Adverse Event Report: RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS CALIFORNIA, INC V60 VENTILATOR; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE Back to Search Results
Model Number V60
Device Problems No Display/Image (1183); Device Emits Odor (1425)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 28jul2020.
 
Event Description
It was reported to philips that the device had an issue with the power board.The customer stated that the device did not display after booting, and reported a burnt like smell.The device was not being used on a patient at the time of the event.There was no adverse event to the patient or user as a result of this event.The device was evaluated by a philips authorized personnel (asp).The asp determined that the motor drive board and power management board are damaged.
 
Manufacturer Narrative
G4: 30sep2020.B4: 30sep2020.The asp stated that the customer declined to have philips repair the device and found a third party to complete the repair on the device.The determination could be made that the device failed to meet specifications.The device was not being used for treatment when the reported event occurred.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
Manufacturer (Section D)
RESPIRONICS CALIFORNIA, INC
2271 cosmos court
carlsbad CA 92011
MDR Report Key10335589
MDR Text Key201554699
Report Number2031642-2020-02536
Device Sequence Number1
Product Code MNT
UDI-Device Identifier00884838025776
UDI-Public(01)00884838025776
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 07/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberV60
Device Catalogue Number1076716
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received07/04/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-