| Model Number |
1588020122 |
| Medical Device Problem Codes |
Poor Quality Image (1408); Device Dislodged or Dislocated (2923)
|
| Health Effect - Clinical Codes |
No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
07/02/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image for about five minutes.
|
| |
|
Event or Problem Description
|
|
It was reported that there was loss of image for about five minutes.
|
| |
|
Additional Manufacturer Narrative
|
|
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: threads worn down on both sides, falls right out during use.Root cause: worn out threads most likely caused by the customer not screwing the coupler to the camera head correctly and/or wear and tear since the coupler has been on the field for two years and has not been repaired.
|
| |
|