MAUDE Adverse Event Report: STRYKER ENDOSCOPY-SAN JOSE PKG, AIM COUPLER, 18MM, C-MOUNT; CONFOCAL OPTICAL IMAGING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER ENDOSCOPY-SAN JOSE PKG, AIM COUPLER, 18MM, C-MOUNT; CONFOCAL OPTICAL IMAGING Back to Search Results
Model Number 1588020122
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Codes No Known Impact Or Consequence To Patient (2692); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 07/02/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was image loss for about five minutes.
 
Event or Problem Description
It was reported that there was image loss for about five minutes.
 
Additional Manufacturer Narrative
The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.Alleged failure: threads worn down on both sides, falls right out during use.Root cause: worn out threads most likely caused by the customer not screwing the coupler to the camera head correctly.Coupler was previously repaired by a third party repair house.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PKG, AIM COUPLER, 18MM, C-MOUNT
Common Device Name
CONFOCAL OPTICAL IMAGING
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key10336134
Report Number0002936485-2020-00311
Device Sequence Number16795044
Product Code OWN
UDI-Device Identifier07613327174359
UDI-Public07613327174359
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K142310
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 09/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1588020122
Device Catalogue Number1588020122
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/10/2020
Initial Date Received by Manufacturer 07/02/2020
Supplement Date Received by Manufacturer07/02/2020
Initial Report FDA Received Date07/28/2020
Supplement Report FDA Received Date09/16/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-