Model Number 90182
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Embolus (1830); Occlusion (1984); Thromboembolism (2654); Embolism/Embolus (4438)
Date of Event 07/03/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The subject device is unavailable to manufacturer.
 
Event or Problem Description
It was reported that during the study procedure with the subject device (retriever), the distal occlusion in aca territory (a1) and new ent (embolization to new territory) to a2 were occurred.The adverse events were treated by recanalization/thrombectomy.The procedure was completed, and the events were resolved with tici (thrombolysis in cerebral infarction score) of 2a was achieved.It¿s indicated that there was relationship between the adverse events to the index stroke, index procedure and the technique was not successful due to subject device (retriever) deployment failure.No other information was provided.
 
Additional Manufacturer Narrative
Based on the results of the dhr review, there is no indication that the device, labeling or packaging failed to meet its specifications when released.The subject device was not available; therefore, a visual inspection as well as a functional evaluation could not be performed.The reported event is covered in the device directions for use (dfu).As well, the risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.Additional information provided by the customer indicated that the technique was not successful due to the stent retriever deployment failure.Since the device was not returned for analysis, the cause of the deployment failure could not be definitively determined.Additionally, based on the information provided, the causal relationship between the subject device and the reported distal occlusion and ent is unknown.Therefore, undeterminable will be assigned to this complaint since review and analysis of all available information fails to indicate an assignable cause or probable assignable cause.
 
Event or Problem Description
It was reported that during the study procedure with the subject device (retriever), the distal occlusion in aca territory (a1) and new ent (embolization to new territory) to a2 were occurred.The adverse events were treated by recanalization/thrombectomy.The procedure was completed, and the events were resolved with tici (thrombolysis in cerebral infarction score) of 2a was achieved.It¿s indicated that there was relationship between the adverse events to the index stroke, index procedure and the technique was not successful due to subject device (retriever) deployment failure.No other information was provided.Updated information: received additional information on 9/4/2020 indicated that the the distal occlusion was caused by iatrogenic embolization during procedure and tortuous anatomy.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TREVO XP PROVUE 4MM X 20MM
Common Device Name
CATHETER, THROMBUS RETRIEVER
Manufacturer (Section D)
STRYKER NEUROVASCULAR-UTAH-SALT LAKE CITY
4870 west 2100 south
salt lake city UT 84120
MDR Report Key10336163
Report Number3012931345-2020-00109
Device Sequence Number19635674
Product Code NRY
UDI-Device Identifier00815742001822
UDI-Public00815742001822
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K132641
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/27/2021
Device Model Number90182
Device Catalogue Number90182
Device Lot Number0000032043
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 07/05/2020
Supplement Date Received by Manufacturer09/08/2020
Initial Report FDA Received Date07/28/2020
Supplement Report FDA Received Date09/30/2020
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other; Required Intervention;
Patient Age63 YR
-
-