| Model Number |
MMT-1715KL |
| Medical Device Problem Code |
Obstruction of Flow (2423)
|
| Health Effect - Clinical Code |
No Consequences Or Impact To Patient (2199)
|
| Date of Event |
07/22/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
|
| |
|
Event or Problem Description
|
|
Information received by medtronic indicated that the insulin pump had insulin flow block alarm.Customer had tried all recommended troubleshooting.No harm requiring medical intervention was reported.The insulin pump will be returned for analysis.
|
| |
|
Additional Manufacturer Narrative
|
|
Device passed the rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test and displacement test.No insulin flow blocked alarm noted during testing.Device functioning properly during testing.(b)(4).
|
| |
|